Biological Indicator Evaluation Resistometer
Product Overview
The Biological Indicator Evaluation Resistometer (BIER) is designed to evaluate the resistance of biological indicators (BIs) and assess chemical indicators (CIs) under controlled sterilization process conditions. It provides a controlled test environment for studying indicator responses to sterilization parameters, supporting sterilization validation laboratories, biological indicator manufacturers, pharmaceutical companies, and medical device manufacturers.
Different sterilization methods require different testing conditions. The product range includes dedicated systems for steam sterilization, hydrogen peroxide sterilization, and ethylene oxide (EO) gas sterilization. Each system is configured for its corresponding sterilization medium, temperature, pressure, and testing procedure.
Available BIER Systems
Steam Sterilization Evaluation Resistometer
The steam BIER system is designed for evaluating biological indicators, chemical indicators, and Bowie-Dick (BD) test products under controlled steam sterilization conditions.
Its working pressure is 0.21 MPa, with a working temperature range of 105–134°C. The system uses a 5 L test chamber and supports rapid steam admission, vacuum operation, and drying procedures.
Hydrogen Peroxide Sterilization Evaluation Resistometer
The hydrogen peroxide BIER system is designed for evaluating biological indicators, chemical indicators, and biological process challenge devices (PCDs) under controlled vaporized hydrogen peroxide conditions.
It operates at 50°C ±0.5°C, with a hydrogen peroxide concentration of 2.7 ±0.3 mg/L. The system uses separate diffusion and test chambers to control hydrogen peroxide vapor generation and exposure conditions.
Ethylene Oxide Gas Sterilization Evaluation Resistometer
The EO BIER system is designed for evaluating biological and chemical indicators under controlled ethylene oxide gas sterilization conditions.
It operates at 54°C ±1°C, with an EO concentration of 600 ±30 mg/L and humidity of 60% ±10%. Automatic humidification and gas concentration monitoring support controlled test conditions.
Technical Specifications
The following specifications are provided separately for each sterilization process.
| Parameter | Steam BIER | Hydrogen Peroxide BIER | EO BIER |
|---|---|---|---|
| Model | BIER-Steam-5 | Hydrogen Peroxide Evaluation Resistometer | EO Evaluation Resistometer |
| Test medium | Saturated steam | Vaporized hydrogen peroxide | Ethylene oxide gas |
| Working temperature | 105–134°C | 50°C ±0.5°C | 54°C ±1°C |
| Working pressure | 0.21 MPa | Pre-injection vacuum: 80 Pa ±5 Pa | Pre-injection vacuum: -95 kPa |
| Test medium concentration | — | 2.7 ±0.3 mg/L | 600 ±30 mg/L |
| Humidity | — | — | 60% ±10% |
| Diffusion / vaporization chamber | — | 135 L | 80 L |
| Test chamber | 5 L | 1.5 L | 1.5 L |
| Medium injection time | Steam admission to set temperature: ≤10 s | ≤5 s | ≤60 s |
| Medium removal time | Drainage: ≤5 s | ≤5 s | ≤60 s |
| Power | 40 kW, including steam generator | 6 kW | 6 kW |
| Dimensions (L × W × H) | 2000 × 1200 × 1880 mm | 900 × 1100 × 1700 mm | 800 × 1020 × 1680 mm |
| Weight | 200 kg | 300 kg | 180 kg |
Note: Specifications are model-specific. The steam system's design pressure is -0.1 to 0.28 MPa, and its design temperature is 150°C. Its vacuum level is below -96 kPa, with a pre-vacuum time of 60 seconds. These are design and operating specifications, not interchangeable ratings.
Testing Functions and Control Systems
Steam Sterilization Testing
The steam system uses PLC control and a color touchscreen interface. Operators can monitor temperature, pressure, time, and alarm status during operation. The system provides process curves, an integrated printer, and USB data storage.
Fixed temperature programs and user-defined test temperatures are available. Its pre-vacuum and post-vacuum drying functions support controlled steam testing procedures.
Hydrogen Peroxide Sterilization Testing
The hydrogen peroxide system uses separate diffusion and test chambers. Its standard testing program includes:
- Preheating of the diffusion and test chambers.
- Vacuum generation in the test chamber.
- Hydrogen peroxide injection and vaporization.
- Adjustment of the test chamber vacuum.
- Hydrogen peroxide injection to the target concentration.
- Exposure for the programmed test period.
- Vacuum removal of residual hydrogen peroxide.
- Air admission and automatic door opening for sample removal.
Standard and enhanced testing programs are available. The enhanced program supports biological PCD performance testing.
Ethylene Oxide Sterilization Testing
The EO system includes automatic humidification and sensor-based monitoring of test conditions. Temperature, humidity, and EO concentration are monitored during operation. Its automatic gas injection and removal functions support controlled exposure procedures.
Configuration and Customization
The BIER systems are available in configurations for steam, hydrogen peroxide, and EO sterilization testing.
Configuration considerations include:
- Sterilization process and corresponding equipment model.
- Test chamber and diffusion chamber arrangement.
- Sample holders and loading accessories.
- Temperature, pressure, humidity, and gas concentration control.
- Test program and operating interface.
- Data recording and process monitoring requirements.
OEM/ODM and project customization services are available. Equipment configuration and special requirements can be discussed according to the intended testing process, laboratory layout, and project specifications.
Manufacturing, Quality Control and Project Support
As a medical and sterilization equipment supplier, Sinicmed Engineering Co., Ltd. provides equipment manufacturing, system design, installation guidance, operator training, and after-sales technical support.
The company provides project-based customization and OEM/ODM services for different equipment requirements. Installation drawings, facility layout support, technical guidance, and spare parts support are available for equipment projects.
Equipment documentation, calibration requirements, and installation or operational qualification support can be discussed according to the selected model and project scope.
Applications
The Biological Indicator Evaluation Resistometer is intended for professional testing and validation environments, including:
- Biological indicator manufacturers evaluating indicator resistance.
- Pharmaceutical manufacturers conducting sterilization process validation.
- Medical device manufacturers assessing sterilization process performance.
- Sterilization validation laboratories conducting controlled indicator testing.
- Testing organizations working with steam, hydrogen peroxide, or EO sterilization processes.