Hydrogen Peroxide Sterilization Evaluation Resistometer
Used to test the resistance of biological indicators (BIs) to vaporized hydrogen peroxide (H2O2) sterilization conditions.
It creates a controlled and reproducible environment that mimics the sterilization process, allowing for accurate and consistent testing of how effectively microbial populations are destroyed by the sterilant.

Hydrogen Peroxide Sterilization Evaluation Resistometer
The Hydrogen Peroxide Sterilization Evaluation Resistometer is specially designed to test biological indicators, chemical indicators, and biochemical PCD performance for hydrogen peroxide gas sterilization.
It is composed of:
● Gasified hydrogen peroxide generation chamber
● Resistance detection (testing) chamber
● Vacuum system
● Hydrogen peroxide purification and gasification injection system
● Electrical control, display, and recording system
● High-precision sensors and connecting pipelines
Both the vaporization and detection chambers, as well as the automatic lift doors, are made from high-strength aluminum alloy for durability and precision.
Test Programs
(Used for biological and chemical indicator testing)
1.Preheat both chambers to the required temperature (e.g. 50 °C)
2.Vacuum to 50 Pa
3.Inject hydrogen peroxide into the generating chamber for vaporization
4.Vacuum the detection chamber to 15 Pa
5.Inject the vaporized H₂O₂ gas to reach the target concentration
6.Maintain sterilization for the set time
7.Vacuum and re-aerate the detection chamber
8.Open the automatic chamber door for sample retrieval
9.Plasma release and air intake reset the generating chamber to zero
(Used for biochemical PCD testing)
1.Preheat the generating chamber to the set temperature (e.g. 50 °C)
2.Vacuum to 50 Pa
3.Inject H₂O₂ into the purification and vaporization system
4.Deliver prepared vapor to the generating chamber
5.Maintain sterilization for the set duration
6.Vacuum, plasma release, and air re-introduction
7.Return the chamber to initial conditions
Technical parameters
|
Parameter |
Specification |
|
Vacuum pressure (before H₂O₂ injection) |
80 Pa ± 5 Pa |
|
Working temperature |
50 °C ± 0.5 °C |
|
H₂O₂ concentration (during operation) |
2.7 ± 0.3 mg/L (real-time display) |
|
Chambers |
Diffusion vaporization chamber – 135 L; Detection chamber – 1.5 L |
|
Door type |
Electric lift doors (both chambers) |
|
H₂O₂ injection devices |
hydrogen peroxide injection devices and vaporization devices |
|
Multiple Pressure transmitters |
Display accuracy 0.1 Pa; Control accuracy ± 5 Pa |
|
Control valve |
Installed between the large chamber and the small chamber to accurately control the sterilization time of the detection chamber. |
|
Temperature control |
Display unit 0.1 °C; Control accuracy ± 0.5 °C |
|
Injection/removal time of H₂O₂ |
≤ 5 s for both |
|
Main control components |
Siemens (Germany) – PLC, touchscreen, temp/pressure modules |
|
Solenoid valves |
SMC (Japan) |
|
Recording system |
Built-in micro-printer for real-time data output |
|
Power |
6 kW |
|
Overall Size (L×W×H) |
900 × 1100 × 1700 mm |
|
Weight |
300 kg |
Hot Tags: hydrogen peroxide sterilization evaluation resistometer, China hydrogen peroxide sterilization evaluation resistometer manufacturers, suppliers, biological indicator evaluation resistometer, Chemical Indicator Evaluation Resistometer, Ethylene Oxide Gas Sterilization Evaluation Resistometer, Ethylene Oxide Indicator Evaluator Resistometer, Hydrogen Peroxide Sterilization Evaluation Resistometer, Steam Sterilization Evaluation Resistometer