Radiopharmaceuticals Sterilizer
video

Radiopharmaceuticals Sterilizer

● Special for radiopharmaceutical vials, saturated steam sterilization
● Auto sliding/lifting door, double safety interlock
● 316L electrolytic polished chamber, durable door seal
● Siemens PLC automatic control, printable sterilization reports
● Customizable programs with power-off memory
● Remote monitoring & intelligent maintenance
● Imported valves & oil-free vacuum pump
● 0.22μm high-efficiency air filter
Send Inquiry
Product Introduction

The SFST series of radiopharmaceutical sterilizers utilize saturated steam as the sterilization medium and are applied in pharmaceuticals and bioengineering, primarily for the sterilization of liquid radiopharmaceuticals packaged in vials.

 

Features

 

Door and Door Seal:

Vertical lifting door (or sliding door), cylinder-driven sliding opening and closing, requiring no manual operation and saving space.

Double-door (single-door customizable) passage type, double-door interlocking, pressurized safety interlock, liquid program and temperature door safety interlock.

Pneumatic pressure driven sealing ring for automatic sealing, new high-quality, high-tear-resistant silicone rubber circular door sealing ring, special coating surface, wear-resistant and smooth, good sealing effect, long service life, up to 1500 cycles.

Inner chamber and door panel material is 316L; if jacketed, jacket material is S30408.

Built-in steam generator material is 316L. The inner chamber, door panel, and evaporator inner surface are mechanically polished and electrolytically polished, with a surface roughness ≤0.4.

Control system uses Siemens configured process automatic control system, S7-1200 series PLC program control, TP series precision panel, enabling real-time monitoring of the sterilization process and sterilization report printing.

The program automatically completes preheating, pulsation, displacement, heating, sterilization, cooling, drying, and exhaust processes. Each process allows for real-time monitoring, recording, and printing of process parameters such as temperature, pressure, time, and operating status.

It includes vial sterilization, BD testing, and vacuum testing programs.

Users can set different sterilization programs as needed. Process parameters can be set and memorized, preventing data and configuration parameter loss in case of power failure.

Features-1

Intelligent Maintenance System: Equipped with the remote monitoring and maintenance module, it can connect to a computer, tablet, or smartphone from anywhere in real-time to access equipment information, including real-time and historical operating data, troubleshooting notifications, and remote software upgrades.

 

Piping System: Stainless steel sanitary piping or PTFE sanitary piping; main control valves use imported German angle seat valves, diaphragm valves, and Japanese SMC solenoid valves, paired with a German Thomas oil-free vacuum pump.

 

Air Filter: Filtration accuracy is 0.22 microns, with a sterilization rate of 99.97%.

 

Product Usage

 

Radiopharmaceuticals

product-596-368
product-596-368

Technical parameters

 

Internal chamber design pressure

-0.1/0.3MPa

Temperature display accuracy

0.1℃

Jacket design pressure

0.3MPa

Pressure display accuracy

0.1KPa

Built-in steam generator (if present)

0.3MPa

Sterilization temperature control accuracy

±02℃

Design temperature

150℃

Temperature uniformity

≤1℃

Maximum operating temperature

139℃

Inner chamber safety valve opening pressure

0.28MPa

Maximum operating pressure

0.25MPa

Jacket safety valve opening pressure

0.28 MPa

 

Radiopharmaceutical Sterilizer Drawing

Hot Tags: radiopharmaceuticals sterilizer, China radiopharmaceuticals sterilizer manufacturers, suppliers, Ampoule Water Shower Sterilizer, P3 Laboratory Biosafety Sterilizer, pharmaceutical sterilizer, Radiopharmaceuticals Sterilizer, Ventilation Sterilizer, Water Bath Sterilizer