P3 Laboratory Biosafety Sterilizer
● Complete Waste Treatment System
● Reliable Biological Isolation
● All-Round Safety Protection
● High-risk Bio Material Sterilization
The biosafety sterilizer is a sterilization device designed and developed in accordance with Biosafety Level 3 requirements. It is equipped with exhaust gas treatment, wastewater treatment, and biological isolation functions, which can fully ensure the safety of operators and the operating environment.
This sterilizer can be widely used for the sterilization of animal carcasses, biological products, instruments, sterile garments, medical dressings, and other items contaminated with highly pathogenic bacteria and viruses.
Design Standards
|
No. |
Content Description |
| 1 |
GB 150.1–GB 150.4 Pressure Vessels |
| 2 |
GB 8599 Technical Requirements for Large Steam Sterilizers (Automatic Control Type) |
| 3 |
GB 4793.1-2007 Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use - Part 1: General Requirements |
| 4 |
GB 4793.4-2001 Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use - Particular Requirements for Autoclaves for the Treatment of Medical Materials in Laboratories |
| 5 |
GB/T 14710 Environmental Requirements and Test Methods for Medical Electrical Equipment |
| 6 |
GB 18278 Sterilization of Health Care Products - Requirements for Validation and Routine Control of Industrial Moist Heat Sterilization |
| 7 |
JB/T 4711 Coating and Transportation Packaging of Pressure Vessels |
| 8 |
YY 0076 Classification and Technical Requirements for Coatings of Metal Parts |
| 9 |
GB/T 191 Packaging - Pictorial Marking for Handling of Goods |
| 10 |
Chinese Pharmacopoeia (2015 Edition) |
| 11 |
TSG 21-2016 Safety Technical Supervision Regulation for Fixed Pressure Vessels |
| 12 |
Relevant regulations and guidelines of the FDA (U.S. Food and Drug Administration) |
| 13 |
Relevant regulations and guidelines of the EMEA (European Medicines Agency) |
| 14 |
GB 50346-2011 Technical Code for Biosafety Laboratory Buildings |
| 15 |
GB 14925 Laboratory Animals - Environment and Facilities |
Product Use:Toxic waste


Performance characteristics
1.The sealing ring is automatically sealed by inflation pressure. A new high-quality, high tear-resistant silicone rubber circular hollow door gasket is used, with a special surface coating for wear resistance and smoothness, providing better sealing performance and a longer service life of up to 1,500 cycles.
2.The main body adopts a rectangular cross-section with a European-style annular reinforcing jacket and a flangeless structure. The sealing groove is directly welded to the annular jackets at both ends (GETINGE structure), ensuring a long fatigue life.
3.Control system: Imported Siemens PLC and color touch screen. The entire process is automatically controlled by programs, with real-time display of operating status, temperature, pressure, time, and other parameters.
4.Traceability system: Equipped with a communication interface for connection to a traceability system, providing software communication protocols for integration.
5.Sterilization programs include textiles, instruments, hollow instruments, rubber, liquids, culture media, waste, and user-defined programs, as well as test programs such as BD test and vacuum leakage test.
6.Built-in micro pin printer for real-time printing of process data; configured with a paperless recorder.
7.Piping system: Stainless steel hygienic piping with clamp connections; all welds are completed by automatic orbital welding machines. Main control valves use imported German angle seat valves and solenoid valves; equipped with a German SPECK direct-coupled water ring vacuum pump.
8.Biosafety system:
● A fully enclosed isolation sealing wall welded to the chamber; all cables and pipelines passing through the sealing wall adopt gas-tight sealing structures, ensuring strict biological isolation for double-door operation.

● Equipped with an online sterilization system for wastewater collection, meeting biosafety sterile discharge standards.

● Equipped with a high-efficiency exhaust steam filter with a filtration accuracy of 0.22 μm and a bacterial removal efficiency of 99.99%, featuring online SIP to ensure exhaust gas is sterilized through the HEPA filter.
● Equipped with a return-air high-efficiency filter with a filtration accuracy of 0.22 μm and a bacterial removal efficiency of 99.99%, featuring online SIP.

9.Equipped with an automatic drain temperature control system; drain temperature < 60 °C.
10.Ensures strict biological isolation for double-door operation.
Technical parameters
|
Chamber design pressure |
-0.1/0.3MPa |
Temperature display accuracy |
0.1℃ |
|
Jacket design pressure |
0.3MPa |
Pressure display accuracy |
0.1KPa |
|
Built-in steam generator (if applicable) |
1.0 MPa |
Sterilization temperature control accuracy |
±0.2℃ |
|
Design temperature |
150℃ |
Temperature uniformity |
±1℃ |
|
Maximum operating temperature |
139℃ |
Negative pressure pulsed vacuum |
Amplitude-80~0KPa, Times 3~9 |
|
Maximum operating pressure |
0.25MPa |
Differential pressure pulsing |
Amplitude-80~80KPa, Times 1~9 |
|
Chamber safety valve set pressure |
0.28MPa |
Positive pressure pulsing |
0~9 Times |
|
Jacket safety valve set pressure |
0.28 MPa |
Ultimate vacuum level |
-96 KPa |
Model
|
Model |
Volume (L) |
Chamber size W*H*D |
Boundary dimension (L*W*H) |
Weight (kg) |
Steam consumption kg/cycle |
Water consumption kg/cycle |
Power |
|
PLBS-0.25 |
250 |
550*550*845 |
1200*1290*1850 |
745 |
16 |
120 |
380V,50Hz,2/20KW |
|
PLBS-0.3 |
300 |
550*550*1000 |
1350*1290*1850 |
820 |
18 |
140 |
380V,50Hz,2/20KW |
|
PLBS-0.36 |
360 |
610*610*1000 |
1350*1350*1850 |
870 |
20 |
160 |
380V,50Hz,2/30KW |
|
PLBS-0.45 |
450 |
652*690*1000 |
1350*1400*1950 |
965 |
22 |
180 |
380V,50Hz,2/30KW |
|
PLBS-0.6 |
600 |
610*910*1200 |
1500*1350*1900 |
1200 |
35 |
320 |
380V,50Hz,3.5/36KW |
|
PLBS-0.8 |
800 |
610*910*1500 |
1800*1350*1900 |
1350 |
47 |
400 |
380V,50Hz,3.5/48KW |
|
PLBS-1.0 |
1000 |
610*910*1800 |
2100*1350*1900 |
1550 |
55 |
500 |
380V,50Hz,3.5/54KW |
|
PLBS-1.2 |
1200 |
680*1180*1500 |
1800*1420*2000 |
1650 |
65 |
600 |
380V,50Hz,3.5/72KW |
|
PLBS-1.5 |
1500 |
680*1180*1850 |
2100*1420*2000 |
1850 |
80 |
750 |
380V,50Hz,4/90KW |
|
PLBS-2.0 |
2000 |
900*1380*1620 |
1970*1750*2200 |
2300 |
106 |
1000 |
380V,50Hz,4.5KW |
|
PLBS-2.5 |
2500 |
900*1380*2050 |
2350*1750*2200 |
2700 |
132 |
1250 |
380V,50Hz,6KW |
|
PLBS-3.0 |
3000 |
900*1380*2420 |
2750*1750*2200 |
3500 |
160 |
1500 |
380V,50Hz,8KW |
|
PLBS-4.0 |
4000 |
900*1380*3250 |
3550*1850*2200 |
4500 |
190 |
1750 |
380V,50Hz,8KW |
Hot Tags: p3 laboratory biosafety sterilizer, China p3 laboratory biosafety sterilizer manufacturers, suppliers, Ampoule Water Shower Sterilizer, Dry Heat Sterilizer, P3 Laboratory Biosafety Sterilizer, pharmaceutical sterilizer, Ventilation Sterilizer, Water Bath Sterilizer